COMPARISON OF THE EFFECTIVENESS OF RADIATION SHIELD WALL BETWEEN LEAD-LAYERS AND PLASTERING BRICK-LAYERS Ardi Soesilo Wibowo; Edy Cahyono; Retno Sri Iswari; Kholik Al Amin; Marichatul Jannah
Jurnal Riset Kesehatan Vol 11, No 2 (2022): NOVEMBER 2022
Publisher : Poltekkes Kemenkes Semarang

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.31983/jrk.v11i2.9307

Abstract

The rate of exposure to X-ray radiation on the radiation shielding wall at the Laboratory 3 of Radiology Study Program Purwokerto Diploma Three Program has been analyzed, to find out the difference in the effectiveness of the radiation shielding wall between a 10 cm thick Lead (2 mm) coated partition and a 28 cm thick stucco brick wall. Measurements were made using a radiation source, namely a mobile unit X-ray machine with a Fluke survey meter radiation measuring instrument. Measurement of the rate of exposure to X-ray radiation is carried out by adjusting the distance of the radiation source with the radiation shield wall from 100 cm, 150 cm and 200 cm and taken from 5 measurement points that represent the radiation shield wall. The measurement results show that measurement point C produces the highest radiation exposure rate and measurement point A produces the lowest radiation exposure rate for Do and D. There is a radiation exposure rate of more than 1 μSv/Hr after passing through a 10 cm thick Lead-coated partition wall (2 mm) at a distance of 100 cm to 150 cm. Radiation shielding walls of walls covered with stucco bricks with a thickness of 28 cm were more effective than partitions covered with lead (2 mm) with a thickness of 10 cm. It is necessary to pay attention to aspects of radiation protection, the use of a mobile unit X-ray machine in placing the X-ray tube in the direction of the X-ray tube and, the radiation source distance of at least 2 meters from the lead-coated partition radiation wall.
IMPROVING BITEWING RADIOGRAPHY: EVALUATING THE EFFECT OF A DENTAL X-RAY POSITIONER ON IMAGE QUALITY IN DENTAL EXAMINATIONS Marichatul Jannah; Ardi Soesilo Wibowo; Agustina Dwi Prastanti; Wingghayarie Patra Gandhi
Jurnal Riset Kesehatan Vol 12, No 1 (2023): MAY 2023
Publisher : Poltekkes Kemenkes Semarang

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.31983/jrk.v12i1.9769

Abstract

Bitewing radiography is a valuable tool in detecting and monitoring dental decay and other oral health problems and is typically performed as part of a routine dental examination. It has certain limitations that should be considered. Some of these limitations may cause discomfort or pain to the patient if not positioned correctly. Some patients may find biting down on the film holder difficult or have a strong gag reflex, making the procedure uncomfortable or even impossible. Bitewing radiography can be costly, especially if it needs to be performed regularly, which may limit access to this diagnostic tool for some patients. This study aimed to evaluate the effect of a dental X-ray positioner with a silicone layer on the quality of radiographic images in posterior bitewing dental examinations. Using this positioner was expected to improve the precision of radiographic interpretation and subsequent patient treatment. The study used a multivariate general linear model to analyze the data obtained from radiographic images using the X-ray positioner with the silicone layer and the conventional X-ray positioner. The results showed no significant difference in image quality between the two positioners, indicating that adding the silicone layer did not significantly improve image quality. However, using any X-ray positioner is still beneficial in ensuring accurate radiographic interpretation and subsequent patient treatment.